<SUM>
<HD SOURCE="HED">SUMMARY:</HD>
The Food and Drug Administration (FDA, Agency, or we) is classifying the device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for the device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test's classification. We are taking this action because we have determined that classifying the device into class II (special controls) will provide a reasonable assurance of safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices.
</SUM>
<EFFDATE>
<HD SOURCE="HED">DATES:</HD>
This order is effective August 16, 2024. The classification was applicable on March 17, 2021.
</EFFDATE>
<FURINF>
<HD SOURCE="HED">FOR FURTHER INFORMATION CONTACT:</HD>
Uwe Scherf, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. 3110, Silver Spring, MD 20993-0002, 301-796-5456,
<E T="03">Uwe.Scherf@fda.hhs.gov.</E>
</FURINF>
<SUPLINF>
<HD SOURCE="HED">SUPPLEMENTARY INFORMATION:</HD>
<HD SOURCE="HD1">I. Background</HD>
Upon request, FDA has classified the device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test as class II (special controls), which we have determined will provide a reasonable assurance of safety and effectiveness. In addition, we believe this action will enhance patients' access to beneficial innovation, in part by placing the device into a lower device class than the automatic class III assignment.
The automatic assignment of class III occurs by operation of law and without any action by FDA, regardless of the level of risk posed by the new device. Any device that was not in commercial distribution before May 28, 1976, is automatically classified as, and remains within, class III and requires premarket approval unless and until FDA takes an action to classify or reclassify the device (see 21 U.S.C. 360c(f)(1)). We refer to these devices as “postamendments devices” because they were not in commercial distribution prior to the date of enactment of the Medical Device Amendments of 1976, which amended the Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to classify or reclassify a device into class I or II. We may issue an order finding a new device to be substantially equivalent under section 513(i) of the FD&C Act (see 21 U.S.C. 360c(i)) to a predicate device that does not require premarket approval. We determine whether a new device is substantially equivalent to a predicate device by means of the procedures for premarket notification under section 510(k) of the FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device through “De Novo” classification, a common name for the process authorized under section 513(f)(2) of the FD&C Act. Section 207 of the Food and Drug Administration Modernization Act of 1997 (Pub. L. 105-115) established the first procedure for De Novo classification. Section 607 of the Food and Drug Administration Safety and Innovation Act (Pub. L. 112-144) modified the De Novo application process by adding a second procedure. A device sponsor may utilize either procedure for De Novo classification.
Under the first procedure, the person submits a 510(k) for a device that has not previously been classified. After receiving an order from FDA classifying the device into class III under section 513(f)(1) of the FD&C Act, the person then requests a classification under section 513(f)(2) of the FD&C Act.
Under the second procedure, rather than first submitting a 510(k) and then a request for classification, if the person determines that there is no legally marketed device upon which to base a determination of substantial equivalence, that person requests a classification under section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA is required to classify the device by written order within 120 days. The classification will be according to the criteria under section 513(a)(1) of the FD&C Act. Although the device was automatically placed within class III, the De Novo classification is considered to be the initial classification of the device.
When FDA classifies a device into class I or II via the De Novo process, the device can serve as a predicate for future devices of that type, including for 510(k)s (see section 513(f)(2)(B)(i) of the FD&C Act). As a result, other device sponsors do not have to submit a De Novo request or premarket approval application to market a substantially equivalent device (see section 513(i) of the FD&C Act, defining “substantial equivalence”). Instead, sponsors can use the less-burdensome 510(k) process, when necessary, to market their device.
<HD SOURCE="HD1">II. De Novo Classification</HD>
On May 19, 2020, FDA received Biofire Diagnostics, LLC's request for De Novo classification of the BioFire Respiratory Panel 2.1 (RP2.1) device. FDA reviewed the request in order to classify the device under the criteria for classification set forth in section 513(a)(1) of the FD&C Act.
We classify devices into class II if general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls that, in combination with the general controls, provide reasonable assurance of the safety and effectiveness of the device for its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the information submitted in the request, we determined that the device can be classified into class II with the establishment of special controls. FDA has determined that these special controls, in addition to the general controls, will provide reasonable assurance of the safety and effectiveness of the device.
Therefore, on March 17, 2021, FDA issued an order to the requester classifying the device into class II. In this final order, FDA is codifying the classification of the device by adding 21 CFR 866.3981.
<SU>1</SU>
<FTREF/>
We have named the generic type of device as device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test, and it is identified as an in vitro diagnostic device intended for the detection and identification of SARS-CoV-2 and other microbial agents when in a multi-target test in human clinical respiratory specimens from patients suspected of respiratory infection who are at risk for exposure or who may have been exposed to these agents. The device is intended to aid in the diagnosis of respiratory infection in conjunction with other clinical, epidemiologic, and laboratory data or other risk factors.
<FTNT>
<SU>1</SU>
FDA notes that the “ACTION” caption for this final order is styled as “Final amendment; final order,” rather than “Final order.” In December 2019, FDA began adding the term “Final amendment” to the “ACTION” caption for these documents, typically styled “Final order,” to indicate an amendment to the Code of Federal Regulations. This editorial change was made in accordance with the Office of Federal Register's (OFR) interpretations of the
<E T="04">Federal Register</E>
Act (44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and parts 21 and 22), and the Document Drafting Handbook.
</FTNT>
FDA has identified the following risks to health associated specifically with this type of device and the measures required to mitigate these risks in table 1.
<GPOTABLE COLS="2" OPTS="L2,i1" CDEF="s50,r150">
<TTITLE>Table 1—Device To Detect and Identify Nucleic Acid Targets in Respiratory Specimens From Microbial Agents That Cause the SARS-CoV-2 Respiratory Infection and Other Microbial Agents When in a Multi-Target Test Risks and Mitigation Measures</TTITLE>
<CHED H="1">Identified risks</CHED>
<CHED H="1">Mitigation measures</CHED>
<ROW>
<ENT I="01">Risk of an inaccurate test result (false positive or false negative result) leading to improper patient management</ENT>
<ENT>
Certain labeling information, including limitations, warnings, device descriptions, explanation of procedures, and performance information identified in special controls (1), (3), (5), and (6);
Use of certain specimen collection devices identified in special control (2);
Certain design verification and validation, documentation of certain analytical studies and clinical studies, risk analysis strategies, and device descriptions identified in special control (4); and Testing of characterized viral samples and labeling information identified in special control (7).
</ENT>
</ROW>
<ROW>
<ENT I="01">Misinterpretation of test results leading to misdiagnosis and associated risk of false test results</ENT>
<ENT>
Certain labeling infor
━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━